NIST’s NIST-4 Kibble balance, already used in work underpinning the kilogram’s 2019 redefinition, received a custom quantum Hall array resistance standard (QHARS). The new arrangement directly supplies a precise resistance reference to the balance, rather than relying on a separately calibrated resistor artifact. NIST said eliminating that calibration step cuts resistance uncertainty and improves measurement accuracy. [story-s1]
The balance determines mass by comparing gravity’s pull with electromagnetic force from current through a coil. NIST’s QHARS uses graphene and 13 smaller elements connected in parallel; together they support greater current than the older gallium-arsenide resistance device, which could not be incorporated directly for the currents needed in the balance. Measurements of 50-gram masses made using multiple QHARS devices closely agreed, according to NIST. [story-s1]
The advance was not a compact, room-temperature instrument: the graphene standard still required cooling to a few degrees above absolute zero and a strong magnetic field. NIST framed room-temperature, low-field operation and programmability as possibilities for future versions. That distinction matters: the demonstrated result was an integrated, more accurate measurement setup, while broader convenience gains remained research goals. Measurement standards matter beyond specialized laboratories because traceable mass measurements underpin routine weighing and comparisons across science and commerce; NIST’s report describes a precision improvement, not a claim that everyday scales changed immediately. [story-s1]
Source documents
- Artifacts, Begone! NIST Improves Its Flagship Device for Measuring Masshttps://www.nist.gov/news-events/news/2023/01/artifacts-begone-nist-improves-its-flagship-device-measuring-mass